Sony SH800

Room: TAC S617

Contact: Zhao.Zhao@yale.edu (203)737-4715 (office)

Technical Specifications

  • Available for self-sorting (after training) of murine samples and human samples that have tested negative for HIV, HEPB and HEPC.
  • Multilaser multiparameter cytometer for cell sorting or analysis
  • Single interrogation point of four lasers (405 nm, 488 nm, 561 nm, and 638 nm)
  • 6 color detection
  • Chip-based design (70 um or 100 um)
  • Intuitive software and advanced automation for set-up, acquisition, sort and analysis
  • Asceptic 2-way sorting

NOTE: Because of the co-linear lasers and single interrogation point, only 1 fluorophore from each channel should be selected. For example APC-Cy7 and PE-Cy7 cannot be distinguished.

Sony SH800

PMT Filter Long Pass Fluorochrome
FL1 455/50 Pacific Blue, BV421, Hoechst, DAPI.CFP
FL2 525/50 487.5 LP BV510, FITC, Alexa Fluor 488, GFP, YFP
FL3 600/60 561 LP BV605, PE, PE-Texas Red, RFP, mCherry
FL4 665/30 639 LP APC, Alexa Fluor 647, PE-Cy5, BV650
FL5 720/60 685 LP Alexa Fluor 700, PerCP-Cy5.5, BV711
FL6 785/60 752 LP PE-Cy7, APC-Cy7, BV785