Sony SH800
Room: TAC S617
Contact: Zhao.Zhao@yale.edu (203)737-4715 (office)
Technical Specifications
- Available for self-sorting (after training) of murine samples and human samples that have tested negative for HIV, HEPB and HEPC.
- Multilaser multiparameter cytometer for cell sorting or analysis
- Single interrogation point of four lasers (405 nm, 488 nm, 561 nm, and 638 nm)
- 6 color detection
- Chip-based design (70 um or 100 um)
- Intuitive software and advanced automation for set-up, acquisition, sort and analysis
- Asceptic 2-way sorting
NOTE: Because of the co-linear lasers and single interrogation point, only 1 fluorophore from each channel should be selected. For example APC-Cy7 and PE-Cy7 cannot be distinguished.
Sony SH800
PMT | Filter | Long Pass | Fluorochrome |
FL1 | 455/50 | Pacific Blue, BV421, Hoechst, DAPI.CFP | |
FL2 | 525/50 | 487.5 LP | BV510, FITC, Alexa Fluor 488, GFP, YFP |
FL3 | 600/60 | 561 LP | BV605, PE, PE-Texas Red, RFP, mCherry |
FL4 | 665/30 | 639 LP | APC, Alexa Fluor 647, PE-Cy5, BV650 |
FL5 | 720/60 | 685 LP | Alexa Fluor 700, PerCP-Cy5.5, BV711 |
FL6 | 785/60 | 752 LP | PE-Cy7, APC-Cy7, BV785 |