TAC Aria-A( S617)
This special order BD FACSAriaII is located in room TAC S617.
The 19-parameter, 5-laser system includes:
- 24 mW 355nm UV laser with 2 PMTs
- 200 mW 405nm Violet laser with 4 PMTs
- 100 mW 488nm Blue laser with 4 PMTs, with two of them dedicated to fluorescence
- 150 mW 532nm Green laser with 5 PMTs
- 140 mW 637nm Red laser with 3 PMTs .
This sorter is equipped with PMT for enhanced detection of the forward scatter.
FACSAria II is capable of sorting with 70, 85 and 100 micron nozzles.
TAC Aria-A will NOT sort:
- Any microbial samples
- Cells derived from the gut with high bacterial content (intestine, pelvic lavage, ...)
- Skin cells
- Cells from the esophagus and upper lung
Self-sorting is not permitted on this machine.
For more detailed information or to schedule a sort, please contact Jennifer.Kelly@yale.edu.
UV Laser - 355 nm, 24 mW
PMT | Bandpass | Longpass | Fluorochrome |
A | 740/35 | 690LP | BUV737 |
B | 450/50 | Indo-1 (Violet), DAPI, Alexa Fluor 350, UV Zombie, Hoechst 33342, or BUV395 with 379/28 BP filter
|
Violet Laser - 405 nm, 200 mW
|
Blue Laser -488nm, 100 mW
PMT |
Bandpass
|
Long Pass
|
|
Fluorochrome |
A | 670/14 | 635LP |
|
PerCP, 7AAD
|
B | 515/20 | 505LP |
|
FITC, AlexaFluor 488, CFSE, GFP, YFP |
C | 485/20 |
|
|
FSC PMT |
|
488/10 |
|
|
SSC
|
Green Laser - 532 nm, 150 mW
|
Red Laser - 637nm, 140 mW
PMT | Bandpass | Long Pass | Fluorochrome |
A | 780/60 | 750LP | APC-Cy7, Alexa Fluor 750, Alexa Fluor 780 |
B | 730/45 | 695LP | APC-Cy5.5, Alexa Fluor 700 |
C | 670/30 | APC, Alexa Fluor 647, iRFP 670 |